Medicines Healthcare Products Regulatory Agency
The SteriFeed colostrum (first milk) collection device should not be used to directly feed babies due to the serious risk of choking if the cap is not removed, the Medicines and Healthcare products Regulatory Agency (MHRA) has warned parents today. The UK regulator urges parents to talk to their healthcare professional for advice on how to feed harvested colostrum to their baby safely.
This follows six reported incidents in which the cap of the SteriFeed device became lodged in the back of a babys mouth when it was not removed before being used to feed the infant directly. No fatalities have been reported, however one infant required emergency surgery.
The device is intended to be used to collect and store colostrum only. It is not intended to be used for feeding but the MHRA is aware that it has sometimes been used in this way.
The issue has been identified with the SteriFeed Colostrum Collector, manufactured by Medicare Colgate, which is used in antenatal colostrum collection by both the public and by healthcare professionals.
Colostrum is often collected from 37 weeks. It is a highly nutritious milk produced to kickstart a newborn babys gut and boost immunity. Colostrum can be frozen and later fed to a baby when required.
The kits are available to buy online for use at home and are provided by some local trusts in supply kits for colostrum collection.
Dr Alison Cave, MHRA Chief Safety Officer, said:
Patient safety is our top priority.
We strongly advise anyone using the SteriFeed Colostrum Collector to contact their healthcare provider for advice on how to collect colostrum and feed an infant safely.
This device is not intended to be used to directly feed your baby due to the risk of choking if the cap is not removed.
Seek medical attention immediately if your child or a child in your care ingests an object and please report any incidences involving these devices to our Yellow Card scheme.
Notes to editors
The MHRA has issued the following advice to the public: * The intended use of this device is for the collection and storage of colostrum only, not to directly feed a baby.* Remove the cap from the top end of the syringe before using the device.* Seek advice from a healthcare professional for guidance on feeding the harvested colostrum to your baby* The manufacturer is currently implementing a design change which would minimise the choking hazard associated with the cap. This change will take approximately six months to complete.* More information can be found on the Device Safety Information page.* For media enquiries, please contact t